PMCC-I Automatic Oil Analysis Equipment Pensky Martens Apparatus Easy Operate

Ask Latest Price
Video Channel
Brand Name :Yuyang
Model Number :PMCC-I
Certification :ISO
Place of Origin :China
MOQ :1 Set
Price :USD2000/Set
Payment Terms :Western Union, T/T, L/C
Supply Ability :100 sets per month
Delivery Time :15 days
Packaging Details :Plywood case
Product name :Automatic Pensky-Martens Closed-Cup Flash Point Tester
Test method :Pensky-Martens Closed-Cup
Test standard :ASTM D93
Operation :Automatic
Color :Grey
Warranty :1 Year
Power supply :AC(220±10%)V, 50Hz.
Size :AC(220±10%)V, 50Hz.
Maximum power consumption :500W
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

PMCC-I Automatic Pensky-Martens Closed-Cup Flash Point Test Equipment

Summary
PMCC-I Automatic Pensky-Martens Closed-Cup Flash Point Tester is designed and manufactured as per standards GB/T 261-2008 Determination of Flash Point - Pensky-Martens closed cup method and ASTM D93 Standard Test Methods for Flash Point by Pensky-Martens Closed Cup Tester. It is used to make determination of the closed cup flash point of the petroleum products.

I. Main technical features
1. Light-touch colored LCD. English man-machine dialog interface. All parameters including preset flash temperature, sample number, atmospheric pressure and date can be set by prompt menu.
2. Simulation tracking of temperature rising curve with time. Prompt function. Test date and time can be modified.
3. 120 groups of data can be saved.
4. Automatically calculate the corrected flash point when the pressure differs from standard atmospheric pressure.
5. Differential detection. Automatic correct the system bias.
6. Automatically open cover, ignite, detect and print result.
7. Safe and easy to operate.


II.technical specifications

1 Power supply AC(220±10%)V, 50Hz.
2 Flash point determination Range: Ambient to 300℃.
Meet requirement of GB/T 261-2008
Meet requirement of GB/T 261-2008
Resulotion: 0.1℃
3 Heating rate Procedure A: (5~6)℃/min
Procedure B: (1~1.5)℃/min
4 Stirring rate Procedure A: (90~120)RPM,
Procedure B: (250±10)RPM
Automatic control and manually adjustable.
5 Igniting mode
Igniting source: gas(or other civilian fuels)Electric ignition.
Gas flame diameter: 3.2mm~4.8mm
6 Ambient temperature ≤ 40℃
7 Relative humidity ≤ 80%
8 Maximum power consumption 500W
9 Overall dimension 410mm×360mm×310mm


III.Accessories

No.
Item
Unit
Qty
1
Oil Cup
Piece
1
2
Power supply wire (250V 6A)
Piece
1
3
Printer paper
Roll
1
4
Burning torch
Piece
1